FT1000 Rack Mounted Electronics
Standard Version

rack electronics
  • High performance displacement measurement system
  • For use against all metallic and non-metallic surfaces, such as steel, aluminium, plastics, glass and ceramics
  • High accuracy and thermal stability
  • Simple calibration procedure
  • May be used with sensor temperatures up to 200°C
  • Internal adjustment for different sensor sizes
  • Standard 1 kHz frequency response, 5kHz optional

DESCRIPTION
FT1000 rack-mount electronics are designed for use with standard CDT sensors and comprise an amplifier and oscillator, which are housed in a single or multi-channel channel ABS enclosure.

The system has the highest performance specification of the current range and is designed to operate with sensor temperatures up to 200°C. Power supply is 230VAC, with 115VAC and 12VDC versions available to special order. System output is 0-10V analogue.



SPECIFICATION


SYSTEM PART NUMBERS
Single Channel System Comprising Amplifier FT1000-RMA
  Oscillator FT1000-OSC
  Enclosure FT1000-ENC


MEASUREMENT RANGE
Sensor Size Electrode Linear range Max Working Typical Capacitance Range Over Linear Range
Ø8.0mm Ø2.0mm 0.10 - 2.0mm 5.0mm 0.280pF to 0.014pF
Ø15.0mm Ø4.0mm 0.20 - 4.0mm 10.0mm 0.560pF to 0.028pF
Ø30.0mm Ø8.0mm 0.40 - 8.0mm 20.0mm 1.120pF to 0.056pF

Stated measurement ranges are against metallic surfaces and may be reduced when applied to non-metallic materials. Figures quoted are for CDT series sensors calibrated with FT1000 rack mount electronics. Sensors should be grounded to earth during normal operation. Larger sensors are available to special order.

GAIN RANGE 100:1 variable gain amplifier with 1/X output
Maximum Target Capacitance 2.5pF full scale
Front Panel Potentiometer Scaling 10 turn pot allows full scale output to be developed from 10% to 110% of maximum target Capacitance
   
"Low" gain setting will set to 2.50pF full scale & measure down to 0.025pF
"High" gain setting will set to 0.25pF full scale & measure down to 0.0025pF



PERFORMANCE
Accuracy ± 0.1% of full scale
Repeatability ± 0.05% of full scale
Stability 100ppm/°C typical
Filter Type 3 pole Butterworth 1kHz (100Hz to 5kHz by resistor change)
Rise Time 300 µs (1kHz filter)
     
INPUTS Sensor Type CDT series (low temp)
  Connector Type Microdot
  Cable Type Low noise 100% screened flexible coaxial
  Cable Length 2 metres standard, 5 metres maximum
  Cable Diameter 2.0mm
OUTPUTS Connector Type BNC
  Cable Type 50 Ohm coaxial
  Cable Length Up to 100 metres
  Range 0-10V analogue
TEMP RANGE Electronics 0 - 50°C
  Sensor 0 - 200°C
POWER SUPPLY Standard Supply 230VAC, 50Hz or 115VAC 60Hz
12VDC version available to special order
GENERAL Module size (W x D x H) 60mm x 250mm x 80mm
  Mounting Stand alone enclosure

 

 